Handbook 44 scale tolerance calculator

Maintenance and acceptance tolerances for Class I, II, III, IIII and III L scales at every test load, with the paragraph each one comes from. Free, no sign-in, and nothing you type leaves this page.

NIST Handbook 44 (2026)

The scale

From the data plate. Results update as you type.

Accuracy class

Tolerance
Sets the shift test in the plan.
Test loads and readings Add the indication to see the error and whether it's within tolerance.
Test loadReading (optional)Remove

Results

Tolerance at each load

Suggested test plan

Loads to use, from HB 44's test notes (N) and Table 4, with each load's tolerance. Round them to the weights you carry.

Class limits

A convenience; Handbook 44 is the authority. Worked out from NIST Handbook 44 (2026), Section 2.20 Scales. Check the current edition and your state's adoption before you rely on a result.

Maintenance or acceptance tolerance?

Acceptance tolerance

The tighter one. It applies to a new or newly reconditioned scale when it's first put in service, and for a short time after it's placed in service or repaired, as HB 44 defines it. It's one-half of the maintenance tolerance (T.N.3.2).

Maintenance tolerance

The one a scale in service is held to on routine tests and inspections. It comes from Table 6: so many divisions (e) of tolerance, more as the test load grows past each band for the class (T.N.3.1).

Divisions and class

Tolerances are counted in the verification division e, which is the displayed division d on Class III, IIII and III L scales. The number of divisions n is capacity ÷ e, and Table 3 sets the range of n each class allows.